Product Description:
P1202-4 50R is a 50 ohm pulse group magnetic field probe manufactured by Langer EMV GMBH, Germany, which is represented by EUTTEST Shenzhen for sales and quotation service.P1202-4 50R probe is compatible with theP1202-4 The rotational characteristics, measurement functions and other characteristics of the probe are the same, the only difference is that the input resistance of this probe has been changed to 50 ohms. The 50R probe has a 50 ohm shorting resistor connected to ground, which makes its current flow 1/2 that of the P1202-4 field source probe without termination resistor.

P1202-4 50R Schematic for Testing IC Immunity to Pulsed Magnetic Fields with Pulse Group Generator
The P1202-4 50R Pulsar Magnetic Field Probe also uses the output voltage signal from the pulsar generator to generate the magnetic field as shown in the figure above. The P1202-4 50R is also available with Fsicher or SHV connectors.
Product technical parameters:
<<<<提醒:左右滑动表格>>>>Probe Model | P1202-4 50R |
manufacturer | Langer EMV-Technik |
agent | Shenzhen EUT TEST |
Input impedance | 50Ω |
Pulse Properties - Shape | 5/50ns |
Pulse Voltage | Maximum ±8kV |
Internal Voltage Probe | Coupling factor: -26dBΩ Test Output Port: 50 Ohm, SMB Shunt: 0.1 ohm |
Pulse Input Connector | 50 Ω Fischer (D103A023) |
size | 180*96*96mm |
Optional Kit | P1202-4 / P1302-4 set Electric transient field coupling source group |
Pulse shape:
The coupled voltage shape curve of the P1202-4 50R probe can be measured using an oscilloscope with a 2kV pulse train EFT signal input, as shown below:

P1202-4 Coupling Voltage Measurement Plot for 50R Probe
The P1202-4 50R and P-1202-4 can output different voltage values when the same EFT voltage signal is input using a pulse train generator, as shown in the figure below:

The P1202-4 50R and P-1202-4 can output different voltage values when the same EFT voltage is input.