DGA 9552N Schwarzbeck 18GHz Fixed Attenuator

Order Number: EUTS0466
Brand: Schwarzbeck

Min. order: 1 Unit
Delivery: Click on the link above to inquire

Standards: Common Test Standards
Industry: General Electronics

Warranty: Original Import ; New Products ; Authorized Sales
Training: On-site training; Telephone; Email
Logistics: Shenzhen to National
Payment: Internet Banking T/T

Product Price:

Introduction:

The DGA 9552N is a Schwarzbeck designed and manufactured product that can be used in the DC-18GHz frequency range and can be configured as a 3dB attenuator, 6dB attenuator, 10dB attenuator, 20dB attenuator, 30dB attenuator, or 40dB attenuator, and you can choose one of the attenuation values, but note that the attenuation values are factory fixed. factory-fixed, you can also order all together, the product is authorized by the Shenzhen EUTTEST agent sales quotes and after-sales service. During calibration of EMI receivers, it is necessary to connect an additional attenuator directly to the output of the pulse generator. The nominal attenuation value is printed on the housing of the attenuator and we calibrate each product individually and provide factory calibration data.

If you need a cheaper attenuator that can only be used up to 2GHz, please shop around:DGA 9553BNC

Test application:

The attenuator DGA 9552N sequence is designed to be bi-directionally usable, so you can change the direction of use of the attenuator at will. There are a wide variety of applications for this attenuator:

  • The measurement range of the spectrum analyzer can be increased.
  • It prevents overloading of any kind of meter; it can be used to control the linearity of meter readings.
  • Measurement uncertainty due to mismatched impedances can be minimized.

Technical parameters:

Product Specification Sheet:

“Reminder: swipe tables left and right”
Product model
DGA 9552N
manufacturer
Schwarzbeck, Germany
China Agent Sales
Shenzhen EUTTEST
Product name
Fixed attenuator
Available frequency
DC- 18 GHz
Selectable attenuation value size
  • 3dB fixed attenuator
  • 6dB fixed attenuator
  • 10dB fixed attenuator
  • 20dB fixed attenuator
  • 30dB fixed attenuator
  • 40dB fixed attenuator
power (output)
  • 5 W / 25 °C, 4 W / 45 °C, 2 W / 4 °C, 1 W / 4 °C, 1 W / 4 °C, 1 W / 4 °C
  • 3 W / 65 °C, 2 W / 85 °C, 2 W / 85 °C, 1 W / 85 °C, 3 W / 65 °C, 1 W / 85 °C
  • 1 W / 105 °C
  • 1 kW peak (5 µsec. pulse width, 0.25% duty cycle)
  • 2 W / 85 °C, 1 W / 105 °C
VSWR
  • 1.15 (DC - 4 GHz)
  • 1.20 (4 - 8 GHz)
  • 1.25 (8 - 12.4 GHz)
  • 1.35 (12.4 - 18 GHz)
Length of different structure types
56.4 mm (type A see below)
From: 45.2 (3 dB, type B)
To: 48.5 mm (40 dB, type B)
(an official) standard
MIL-DTL-3933
impedance
50Ω
“Reminder: swipe tables left and right”

Graph of related technical parameters:

The DGA 9552N attenuator is designed in both TypeA and TypeB form factors to account for heat dissipation and test area constraints, and both form factors are identical in terms of technical parameters such as S21.

The DGA 9552N attenuator is available in two configurations (type A and type B).

The attenuator is available in two configurations (type A and type B)

S21 Parameter Plot for DGA 9552N with 10dB Option

S21 parameter curve with 10dB option

As mentioned above, the 9552N can also be optionally configured with a variety of attenuation sizes such as 3dB, 6dB, 10dB, 20dB, 30dB, 40dB, etc. For more S21 curve diagrams of the 9552N attenuator configurations, please contact us for more information.

 

Schwarzbeck Official Authorized Dealer:

EUTTEST is an official authorized distributor of SCHWARZBECK, whose full official name is: SCHWARZBECK MESS - ELEKTRONIK OHG

Ltd. is authorized by schwarzbeck to sell all products in China and provide technical service for the products sold.

Entered. 0/100

Sales Hotline:13425158926

Sales mailbox:sales@euttest.com

Sales WeChat account: euttest

EUTTEST-inquiry-wechat Shenzhen EUTTEST Technology Co.

Contact Sales by WeChat

Sales QQ:3127609248

[761] DISCLAIMER: Reproduction and redistribution by anyone without written permission from EUTTEST is prohibited.

Product parameters are subject to the manufacturer's final version.

If you like our products, please press Ctrl + D shortcut to favorite this product.