INS-0805 Circuit Module Output Voltage Pulse 1.2/50 μs, 8 kV, 0.5 J, 500 Ω

Order Number: EUTS0026

Min. order: 1 Unit
Stock: 0
Delivery: 15 weeks

Standards: Common Test Standards
Industry: General Electronics

Warranty: Original Import ; New Products ; Authorized Sales
Training: On-site training; Telephone; Email
Logistics: Shenzhen to National
Payment: Internet Banking T/T

Product Price:

Product Description:

The INS-0805 is an optional circuit module for the INS Sequential Pulse Insulation Test Generator manufactured by EMCpartner, Switzerland and distributed by EUTTEST, with customized outputs.Voltage pulse 1.2/50 μs, 0.5 J, 500 Ω, 8 kV for testing the insulation of samples against voltage pulses.

Technical parameters:

“Reminder: swipe tables left and right”
Equipment Model
INS-0805
manufacturer
EMCpartner
China Agent Sales
EUTTEST
Meet the standard
iec 61180, iec 60255-27, iec 62052-11.
iec 60834-1/2, en 50470-1
Voltage test level 0.5 J
0.5 kV, 1 kV, 1.5 kV, 2 kV, 2.5 kV, 3 kV, 4 kV.
5 kV, 6 kV, 8 kV + 0 % / - 10 %
Pulse capacitance
250nF ±20%
polarities
Positive, negative, alternating
Output impedance
500Ω±10%
Pulse Rise Time
1.2μs ±30%
Pulse duration
50μs ±20%
pulse energy (physics)
0.5 J ±10%
repetition rate
1s - 655s
INS-0805 Control Attributes
operating system
EMCPARTNER FIRMWARE
multilingualism
10 types available
user interface
7-inch touch screen
communications port
Network Port, USB, RS485
BNC port to monitor surge voltage
1V=4KV, accuracy 3%
Pulse voltage display
0.5kV- 8kV, Accuracy 3%
peak detection function
usability
trigger output
BNC, MAX 12V
Trigger Input
Automatic, Manual, External BNC
pulse counter
Programmable 65535
emergency stop
Emergency stop button, BNC input (EUT disabled)
weight
1.5kg
size
24*10*8.5cm
“Reminder: swipe tables left and right”

Contact EUTTEST for detailed parameters of the above options.

Test Configuration Description:

INS-0805 Match requiredINS Sequential Insulation Pulse GeneratorUse:

The INS is a serial model that includes the INS-1250-7K5, INS-1250-15K, INS-1250-22K5, and INS-1250-30K;

Generator:

  • INS-1250-7K5 The following is an example of a customized output for a 7.5 kV impulse 1.2/50 μs, 40 Ω and/or custom output.
  • INS-1250-15K The following is an example of the type of power supply that can be used with a 15 kV impulse 1.2/50 μs, 40 Ω and/or custom output.
  • INS-1250-22K5 The following is an example of a customized output for a 22.5 kV impulse 1.2/50 μs, 40 Ω and/or custom output.
  • INS-1250-30K The following is an example of the type of power supply that can be used with a 30 kV impulse 1.2/50 μs, 40 Ω and/or custom output.

circuit module

INS-0805, 500 ohm output/ 0.5 J with 0.5 kV, 1 kV, 1.5 kV, 2 kV, 2.5 kV, 3 kV,4 kV, 5 kV, 6 kV and 8 kV test levels (this page)

Matching Networks:

CN-INS-500: 500 Ω, or 40 Ω direct output from the generator.

CN-INS-X: 40Ω-4kΩ customizable

Other optional modules:

  • OPT-INS-MEAS: Voltage and Current Test Boards
  • V-PROBE-30K: Voltage Probe
  • EXT-INS-CON: test chamber
  • TC-ST: Safety circuits
  • TC-ADAPTER: Conversion Circuitry
  • WARNING-LAMP: Hazard warning lamps
  • EMERGENCY-STOP: Emergency stop button
  • TEMA3000: Test Software
  • INS serial products have replacedMIG Serial ProductsMultiple models, contact us for more information.
  • Reference: Digital Isolators - High Voltage Enhanced Isolation: Safety Regulations, Standard Definitions and Test Methods

 

Order No. EUTS0026

Entered. 0/100

Sales Hotline:13425158926

Sales mailbox:sales@euttest.com

Sales WeChat account: euttest

EUTTEST-inquiry-wechat Shenzhen EUTTEST Technology Co.

Contact Sales by WeChat

Sales QQ:3127609248

[860] DISCLAIMER: Reproduction and redistribution by anyone without written permission from EUTTEST is prohibited.

Product parameters are subject to the manufacturer's final version.

If you like our products, please press Ctrl + D shortcut to favorite this product.